IEC 60587 6KV High Voltage Tracking and Erosion Resistance Test Apparatus
Specifications:-
Temperature:-23 ± 2 ℃
Dimension:-L1245 * W723 * H2295 mm
Model Number:-D400
Applied voltage:-AC/DC 100V ~ 6000 V±5% adjustable
Material:-SUS 304 Stainless Steel
Standard:-IEC 60587
Drip interval time:-0 – 99.99s continuously adjustable
Test Material:-PE, PVC and SRPVC
Air pressure:-86 ~ 106 kappa
Power supply:-AC220 (±10%) V / 50 HZ
Product Description
IEC 60587 High Voltage Evaluating Resistance to Tracking Test Equipment for Tracking and Erosion
Main Technical Parameters:- | |
Applied voltage | AC/DC 100V ~ 6000 V±5% adjustable |
Applied current | AC/0 ~ 100mA±1% + 5d, DC/0 ~ 100 mA±1% + 3d |
Output voltage drop of the test device | Should not is of exceed 5% when measured at 60 mA current continuous 0.5s. |
Electrode | The thickness of the which is 304 stainless steel upper and lower electrodes is 0.5 mm |
Distance | The distance between the upper and lower electrodes is 50.0 mm±0.5 mm. |
Drip device | Drip device contaminated liquid flow adjusted at 0.075 ml/min – 0.9 ml/min, height can be adjusted, the rate can be testing time can be predicted. |
Drip interval time | 0 – 99.99s continuously adjustable |
Contaminated liquid | Ammonium chloride, a non-ionic wetting agent with the distilled mixture. The resistivity, isooctyl phenyl polyethoxyethano of the contaminated water or deionized water solution at (23±1) centigrade should be (3.95±0.05) Omegala. |
Power supply | AC220 (±10%) V / 50 HZ |
Instrument dimensions | L1245 x W723 x H2295 (mm) |
Frequently Asked Questions:-
What is Erosion Resistance Test?
Sandblasting rig, candidate erosion impact material samples to the function of the particle to ASTM G76-13 determine the capability of performing testing of characteristics according to resistance as an Impact velocity up to 150 m/s
What is 6KV High Voltage?
It measures the input impedance capability of Intended for low energy an AC/DC voltmeter to 6000 V peak designed to extend the voltage AC or DC. The probe provides high accuracy when Overvoltage Category I is used with a voltmeter having 10 MΩ. Applications
What’s considered high voltage?
Define high voltage as, and the National Electrical Manufacturer’s at least 1500 V for voltage as over 100 to direct current. In the United States, above 1000 V for alternating current Association (NEMA) defines as high 230 kV.
IEC 60587 High Voltage Tracking and Erosion Resistance Test Apparatus
You can also check out our Machinery Products.