High Voltage Tracking Test Equipment is suitable for testing the resistance to electric tracing and corrosion of electrical and electronic products, household appliances and its materials. Its standard is in line with GB/T6553-2003 and IEC60587-1984.
Specifications:-
Drip interval time:-0 – 99.99s continuously adjustable
Applied voltage:-AC/DC 100V ~ 6000 V±5% adjustable
Model Number:-D360
Standard:-GB/T6553-2003 and IEC60587-1984.
Test Material:-PE, PVC and SRPVC
Power supply:-AC220 (±10%) V / 50 HZ
Air pressure:-86 ~ 106 kpa
Material:-SUS 304 Stainless Steel
Dimension:-L1245 * W723 * H2295 mm
Temperature:-23 ± 2 ℃
Main Technical Parameters | |
Applied voltage | AC/DC 100V ~ 6000 V±5% adjustable |
Applied current | AC/0 ~ 100mA±1% + 5d, DC/0 ~ 100 mA±1% + 3d |
Output voltage drop of the test device | Should not exceed 5% when the current is measured at 60 mA of continuous 0.5s. |
Electrode | The thickness of the upper and lower electrodes is 0.5 mm, which is 304 stainless steel. |
Distance | The distance between the upper and lower electrodes is 50.0 mm±0.5 mm. |
Drip device | Adjusted, contaminated liquid flow Drip device height can be rate – 0.9 ml/min, test time can be adjusted at 0.075 ml/min predicted. |
Drip interval time | 0 – 99.99s continuously adjustable |
Contaminated liquid | Ammonium chloride, a non-ionic wetting agent with isooctyl phenoxy polyethoxyethanol distilled. The resistivity of the contaminated water or deionized water mixture solution at (23±1) centigrade should be (3.95±0.05) Omega.m. |
Power supply | AC220 (±10%) V / 50 HZ |
Instrument dimensions | L1245 x W723 x H2295 (mm) |
Frequently Asked Questions:-
What is tracking in insulation?
Tracking is the surface of an insulating material wherein an electrical breakdown on the initial exposure is more conductive than more heat, and eventually the original insulator, producing to heat chars the material, and the char more current flow, complete failure.
What causes electrical tracking?
The effect discharge and a column of light created from causes intense heat at the point of the partially ionized gas. The name fact that the ionized gas temperature arc tracking comes from the rises and the electric current remains to produce electrode ends, producing connected to the current an arc shape.
What is IEC 60587?
It describes two of electrical insulating materials for a liquid contaminant and inclined use test methods for the evaluation under severe (45 Hz to 65 Hz) by measurement of the resistance to tracking ambient conditions at power frequencies and erosion, using plane specimens.